dc.description.abstractThe main goal of this project was to design and test a set of collimators for use in measuring x-ray attenuation. Two collimators: a primary collimator and a secondary collimator, were built. The primary collimator was built using 10.0 cm of lead, followed by 7.0 cm of tin, 1.5 mm of copper and 0.5 mm of aluminum, in the respective order. A 1.5 mm thick, copper filter was also used on the aperture of the primary collimator. The secondary collimator was built using 2.5 cm of lead. The cross-sectional size of both the collimators was made to be 15 x 15 cm2. The shape of the aperture in each collimator was made to be 1.5 cm x 3.0 cm. Various features of the collimators were tested in the Varian TrueBeam and Varian Trilogy linear accelerators. The primary collimator was found to best minimize the secondaries produced by the various components inside the gantry and the collimator itself that reach the absorber when placed at least 30.0 cm away from the absorber upstream the beam. The copper filter in the primary collimator was found unnecessary. The secondary collimator was found to best minimize the secondaries produced by the absorber that reach the detector, when placedat least 20.0 cm away from the detector upstream the beam. With both the collimators placed at their optimal positions, the cross-sectional size of the collimated beam measured around the detector, placed at 200 cm from the source, was found to be approximately equal to 1.5 cm x 3.0 cm. The cross-sectional size of the collimated beam at the position of absorber was found to be 1.9 cm x 4.3 cm and that of attenuated x-ray beam was found to be 3.0 x 5.5 cm.
dc.formatapplication/pdf
dc.languageen_US
dc.rightsCopyright is held by the author who has granted the Oklahoma State University Library the non-exclusive right to share this material in its institutional repository. Contact Digital Library Services at lib-dls@okstate.edu or 405-744-9161 for the permission policy on the use, reproduction or distribution of this material.
dc.titleDesign and Testing of Collimators for Use in Measuring X-ray Attenuation
dc.contributor.committeeMemberYukihara, Eduardo G.
dc.contributor.committeeMemberBorunda, Mario F.
osu.filenamePANTHI_okstate_0664M_15531.pdf
osu.accesstypeOpen Access
dc.description.departmentPhysics
dc.type.genreThesis
dc.type.materialtext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record