Automated Rapid Particle Investigation Using Scanning Electron Microscopy
Abstract
The chemical composition of fly ash particles has been known to vary significantly depending on a number of factors. Current bulk methods of investigation including X-Ray Fluorescence and X-Ray Diffraction are thought to be inadequate in determining the performance of fly ash in concrete. It is the goal of this research to develop a method of Automated Rapid Particle Investigation that will not look at fly ash as a bulk material but as individual particles. By examining each particle individually scientists and engineers will have the ability to study the variation in chemical composition by comparing the chemistry present in each particle. The method of investigation developed by this research provides a practical technique that will allow the automated chemical analysis of hundreds, or even thousands, of fly ash particles in a matter of minutes upon completion of sample preparation and automated scanning electron microscope (ASEM) scanning. This research does not examine the significance of the chemical compounds discovered; rather, only the investigation methodology is discussed. Further research will be done to examine the importance of the chemistry discovered with this automated rapid particle investigation technique.
Collections
- OSU Theses [15752]